KANE 455 CPA1 FLUE GAS ANALYSER

GAS SAFE REGISTER TM1

KANE 455 flue gas analyser uses infra-red technology to directly measure CO2 and has an easy to use rotary switch allowing the user to view the test readings taken simply. Multi-fuel : natural gas, propane, butane, LPG, light oils (28/35 seconds) and wood pellets. Features long life CO sensors (5 years), long life infra-red based CO2 sensor (5 years), high accuracy manometer and differential thermometer. Separate reports for : combustion, pressure, let-by and tightness, temperature and CO build up. Comes complete with protective rubber sleeve with magnet for ‘hands free’ use. Supplied with KMCP2 flue probe, 4 x NiMH rechargeable batteries, charger, 2 x pressure fittings, instruction manual and calibration certificate. Independently approved to BS EN 50379 : 2004.

SPECIFICATIONS :
Parameter Range Resolution Accuracy
TEMPERATURE MEASUREMENT
Flue temperature : 0 - 600°C
(0 - 1200 with special probe)
0.1°C ±1.5°C, or ±1% reading
Inlet temperature
Internal sensor :
0 - 50°C 0.1°C ±1.0°C
External sensor : 0 - 600°C 0.1°C ±1.5°C, or ±1% reading
GAS MEASUREMENT
Oxygen : 0 - 21% 0.1°C ±0.3%
Carbon monoxide : 0 - 2000 ppm nom.
4000 ppm max.
for 15 mins.
1 ppm ±5 ppm < 100 ppm,
±5% reading > 100 ppm
Carbon dioxide : 0 - 20% 0.1 % ±0.2%
Efficiency* : 0 - 99.9% 0.1 %
Excess air* : 0 - 250% 0.1 %
CO/CO2 ratio* : 0 - 0.9999 0.0001
DIFFERENTIAL PRESSURE
Normal range ±80 mbar (maximum over range without damage to sensor is ±400 mbar)
±0.2 mbar 0.001 mbar ±0.005 mbar
±1 mbar to 24.999 bar ±0.03 mbar
±80 mbar then 0.01 mbar ±3% of reading
PRE-PROGRAMMED FUELS : Natural Gas, propane, butane, LPG, light oils (28/35 sec), wood pellets.
DIMENSIONS :
Weight : 0.8 kg (1.8 lb), handset with protective cover.
Handset : 200 mm (7.9") x 45 mm (1.8") x 90 mm (3.5").
Probe : L 300 mm (11.8") (including handle) x dia. 6 mm (0.25")
with 240 mm (9.4") long stainless steel shaft, type K thermocouple and 3 m (9 ft) long neoprene hose.
* Calculated
CONTENTS :
455 infra-red flue gas analyser, KMCP2 probe, 4 x NiMH rechargeable batteries and battery charger, 2 x pressure fittings, 1 x pressure hose, 1 x ASP2 appliance sampling probe kit, KMIRP infra red printer, GLD450 battery operated gas leak detector, 1 x KAL1 air/liquid temperature probe, 2 x KPCP clamp-on pipe probes manual, and carrying case.
(Product images are representative and may differ from the actual products supplied.)
Part No.DescriptionUnit PriceQty

Kane ASP2 Appliance Sample Probe Kit

The KANE ASP2 appliance sampling probe kit enables the investigation of fumes/products of combustion from appliances such as cookers, grills and other flueless appliances, as described in BS 7967: 2005. Each of the three probes can be fitted onto the handle assembly using the compression fitting supplied and then connected to the flue gas analyser with the hose supplied. The kit is compatible with most electronic combustion analysers.

KIT CONTENTS :
Grill probe with oval supports and 7 sampling holes over a 250 mm sampling length. Right angled multihole probe with 5 sampling holes over a 125 mm sampling length. Straight open ended probe, 325 mm long. 2 metre hose with handle and compression fitting.
Part No.DescriptionUnit PriceQty